AT-2040 Application Guide

How to Create an Automatic Sensor Test on the AT-2040

The AT-2040 Gen 2 lets you create automatic sensor test setups directly on the instrument. Define the sensor type, sensitivity, reference point, frequency sweep, and amplitude levels, then run a repeatable calibration sequence without building the test by hand each time.

Automatic test workflow

  • Create or select a sensor setup
  • Enter manufacturer and part number
  • Select sensor type and units
  • Define reference frequency and amplitude
  • Add frequency/amplitude test points
  • Run the test and export PDF/CSV results
Automatic sensor testing

What an automatic test does

An automatic sensor test runs a stored set of vibration points and records the measured sensor response at each point. Instead of manually changing the frequency and amplitude for every step, the AT-2040 automatically moves through the predefined sequence.

Automatic tests are useful for full calibrations, repeatable frequency response checks, production verification, and calibration lab workflows where the same sensor model must be tested the same way every time.

Automatic test setup includes

Sensor identity Manufacturer and part number
Sensor setup Type, units, sensitivity
Reference point Reference frequency and amplitude
Test sweep Frequency and amplitude table
Results Graph, table, PDF, CSV
When to use automatic testing

Use automatic tests for repeatable calibration work

Manual mode is best for quick checks and troubleshooting. Automatic Sensor Test mode is best when the same procedure needs to be run consistently across multiple sensors or across multiple calibration events.

01

Full calibration

Run a multi-point calibration sweep and record sensitivity across the sensor’s working range.

02

Production testing

Use the same stored setup for each sensor model so every unit is tested the same way.

03

Incoming inspection

Verify sensors before they are placed into service or added to inventory.

04

Repeatable records

Save test results and export PDF or CSV files for customer documentation and internal quality systems.

Before creating the setup

Plan the sensor test first

Before entering a new sensor setup, review the sensor datasheet, prior certificate, customer requirements, and AT-2040 operating limits. A good automatic test starts with a realistic frequency and amplitude plan.

01

Start with the reference point

Define the main reference frequency and amplitude first. For many accelerometers, this is commonly 100 Hz at 1 g, unless the customer or sensor certificate specifies something different.

02

Choose the sweep points

Select frequencies that cover the sensor’s operating range and your required calibration points. Do not add unnecessary points that do not support the purpose of the test.

03

Check amplitude limits

Make sure each frequency and amplitude combination is within the AT-2040 shaker capability and does not exceed the sensor’s input limit.

Amplitude planning

Use safe frequency and amplitude combinations

The AT-2040 validates frequency and amplitude entries to help protect the shaker and sensor. Lower frequencies are often displacement-limited, while higher frequencies can become force or payload limited.

When planning a custom automatic test, use lower amplitudes when needed and avoid test points that exceed the shaker limit for the sensor mass and fixture mass.

Practical planning guide

  • Use reduced amplitudes at low frequency.
  • Use normal reference amplitudes in the mid-frequency range where practical.
  • Reduce amplitude at high frequency when needed.
  • When in doubt, use a lower vibration level.
  • Do not exceed the sensor’s rated input level.
Create the setup

Create a custom automatic sensor test

Use the Test Setup screen when the sensor is not already in the AT-2040 database or when you need a custom calibration procedure.

01

Open Settings

From the main menu, select Settings.

02

Open Test Setup

Select Test Setup to create or edit a sensor configuration.

03

Enter manufacturer

Enter the sensor manufacturer name. This field is required.

04

Enter part number

Enter the sensor model or part number. This field is required.

05

Select sensor type

Choose the correct input type, such as IEPE, Charge, Voltage, Proximity, Velocity, or 4–20 mA.

06

Save the setup

After entering the setup details and test points, save the configuration to the internal database.

Setup fields

Recommended fields to enter

The exact fields depend on the sensor type, but most automatic tests require sensor identity, sensitivity, units, reference point, and test point data.

Field What to enter Why it matters
Manufacturer Sensor manufacturer name Used to organize the sensor in the database.
Part number Sensor model or part number Used to identify the exact sensor setup.
Sensor type IEPE, Charge, Voltage, Prox, Velocity, 4–20 mA, or other supported type Selects the correct input path and measurement behavior.
Units g, m/s², IPS, mm/s, mils, mm, µm, or transmitter units Controls how the test points and results are displayed.
Sensitivity Rated sensor sensitivity from the datasheet or prior certificate Used to calculate deviation and compare response.
Reference frequency Main calibration reference frequency Used as the baseline for sensitivity and deviation.
Reference amplitude Main calibration vibration level Defines the primary test condition.
Frequency/amplitude table Each automatic test point Defines the sweep the AT-2040 will run automatically.
Sensor type examples

Choose the correct sensor type

The sensor type determines how the AT-2040 routes the input signal, displays units, and calculates sensitivity.

IEPE

IEPE accelerometer

Use for powered IEPE accelerometers, typically measured in mV/g with bias voltage readback.

Charge

Charge accelerometer

Use for charge output sensors, typically measured in pC/g through the charge input path.

Voltage

Voltage output sensor

Use for voltage accelerometers or DC/AC voltage-output sensors, commonly measured in mV/g or V/g.

Velocity

Velocity sensor

Use velocity units such as IPS or mm/s and sensitivity such as mV/IPS or mV/mm/s.

Prox

Proximity probe

Use displacement units such as mils, mm, or µm and monitor gap voltage where applicable.

4–20

4–20 mA transmitter

Use for current-loop vibration transmitters and verify output against the configured full-scale range.

Test point example

Example automatic accelerometer test points

These example points are only a starting point. Your actual test setup should follow your lab procedure, customer requirements, sensor datasheet, and the AT-2040 amplitude limits.

Step Frequency Amplitude Purpose
1 10 Hz Reduced amplitude Low-frequency response check
2 50 Hz Customer-defined Low/mid-frequency response
3 100 Hz Reference amplitude Main sensitivity reference point
4 200 Hz Customer-defined Mid-frequency response
5 500 Hz Customer-defined Frequency response check
6 1,000 Hz Customer-defined Upper frequency response check
Run the test

Run the automatic sensor test

Once the setup is saved, the sensor can be selected from the Sensor Test screen and run automatically.

01

Mount the sensor

Mount the sensor to the AT-2040 using the correct stud, adapter, or fixture.

02

Connect the sensor

Connect the sensor to the correct input channel or adapter cable for that sensor type.

03

Open Sensor Test

From the main menu, select Sensor Test.

04

Select manufacturer

Select the manufacturer from the database list.

05

Select sensor model

Select the saved sensor setup or model number.

06

Start the test

Press Start and let the AT-2040 run through the programmed sequence.

Review results

Use graph view or table view

Automatic test data is shown in graph view by default. The graph shows deviation relative to the reference sensitivity, while table view shows each test point and fills in the results as the test progresses.

During the test, the screen can show status, reference measurement, current step, serial number, graph/table toggle, and Start/Stop control. For IEPE sensors, bias voltage may also be displayed. For proximity probes, gap voltage may be displayed.

Status meanings

  • Ready: test has not started.
  • Testing: automatic sweep is running.
  • Stopped: test was stopped before completion.
  • Complete: all programmed test points are finished.
  • Done: table/step sequence is complete.
Reports

Save and export automatic test results

At the end of the automatic test, the AT-2040 saves the data to onboard memory. The results can be recalled later and exported for customer records or internal quality documentation.

PDF

Calibration certificate

Export a PDF report for documentation, review, or customer delivery.

CSV

Measured data

Export CSV data for spreadsheets, databases, or quality-system records.

USB

File transfer

Use the Reports screen and USB export workflow to move completed test data off the instrument.

Important calibration note

Automatic test does not force a pass/fail decision

The AT-2040 provides measured sensitivity, deviation, phase data where applicable, status, and calibration records. It does not force a universal pass/fail result because each laboratory may have its own tolerance, uncertainty budget, and decision rule.

If a statement of conformity is required, apply your lab’s ISO 17025 decision rule or the specific decision rule requested by the customer.

Troubleshooting

Common automatic test setup problems

Sensor does not appear in the list

Confirm the custom sensor setup was saved, the manufacturer and part number were entered, and the correct manufacturer is selected from the Sensor Test screen.

Amplitude entry is rejected

The selected amplitude may exceed the limit at that frequency or in that unit. Reduce amplitude, change the frequency, or review the shaker limit chart before saving the point.

Results look wrong after changing units

Stay in one unit system while creating the setup. If you change units after entering points, review the point values and re-enter them if needed.

Best practices

Tips for better automatic tests

Use clear naming

Use manufacturer and part number names that are easy for technicians to recognize.

Match the datasheet

Use sensitivity, units, and frequency range that match the sensor documentation.

Do not overload the shaker

Check sensor mass, fixture mass, frequency, and amplitude before creating high-level points.

Review the report

Verify serial number, test points, and results before sending the certificate to the customer.

Need help building automatic tests?

Agate Technology can help you create the right AT-2040 sensor test workflow.

Whether you are building a calibration procedure for IEPE sensors, charge accelerometers, velocity sensors, proximity probes, voltage sensors, or 4–20 mA transmitters, the AT-2040 can support repeatable automatic testing and PDF/CSV result export.

Agate Tech logo

All rights reserved @2026

Follow Us On

Scroll to Top